Inficon LDS3000 Modular Leak Detector Interface descriptio Instrukcja Użytkownika Strona 50

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 56
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 49
50 Fieldbus Communication
Information The PROFIBUS-DP protocol is subject to change. If you are using this
protocol, please ask INFICON for an update.
6.3.2 Read Process Data (Leak Detector PLC)
This data is sent periodically from the leak detector to the programmable logic cont-
roller:
2
0
Gas ballast
Transition 0 -> 1: 0x01 = Gasballast on
Transition 1 -> 0: 0x00 = Gasballast off (if
Gasballast mode != GASBALLAST_ON)
Gasballast
1
2
Zero mode
0=normal
0x04 = 1 ... 2 dec.
0x08 = 2 ... 3 dec.
0x0C = 19/20 part of the value
3
4
CAL mode
0 = external CAL
0x10 = dyn. CAL
0x20…0x30 = not used
Select dyn/
norm
5
6
Sniff/Vac
0=Vacuum
0x40 = Sniff
0x80 = according to PLC-Input
0xC0 = not used
Sniff
*CONFIG:M
ODE
401
7
Byte Bit Name Meaning
Similar to
PLC Output
Similar to
RS232 ASCII
cmd.
Similar to
RS232 LD
cmd.
1
0 (not used) always 1
1 Zero active
0=off
0x02 = on
ZERO active
*STATUS:ZE
RO?
2Error
0=no error
0x04 = error
Error
Status
word
3 Warning
0=no warning
0x08 = warning
Warning
Status
word
4
State
internal
calibration
0=inactive
0x10 = active
0x20/0x30 = not used
CAL active
*STATUS:C
AL?
260
5
6
State
external
calibration
0=inactive
1 = 0x40 = active
2 = 0x80 = waiting for test leak closed
3 = 0xC0 = not used
CAL active
*STATUS:C
AL?
260
7
Przeglądanie stron 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 56

Komentarze do niniejszej Instrukcji

Brak uwag